Gamma dose rate, activity & shielding calculator

Dose rate from a point source at any distance, through lead, steel or concrete — or solve for activity, distance or shield thickness.

Calculator

Calculated — choose the unit for the result.

Calculated — choose the unit for the result.

When solving for shield thickness this is the target dose rate behind the shield.

Calculated — choose the unit for the result.

Results

Dose rate
330 mR/h
Exposure rate
330 mR/h
Air-kerma rate
2.892 mGy/h
Dose-equivalent rate (≈, 1 R ≈ 1 rem)
3.3 mSv/h · 330 mrem/h
Activity
1 Ci · 37 GBq
Distance
1 m · 3.281 ft
Γ used
0.33 R·m²/(Ci·h)
  • Cs-137: Includes the 662 keV gamma of its Ba-137m daughter (85.1 % per decay) in secular equilibrium.
  • Point-source estimate: it applies only when the distance, measured from the center of the source, is at least about three times the largest dimension of the source and its holder. Closer than that, or for a large or spread-out source, the formula does not give a valid figure. Scatter from walls and floors and shielding by the capsule are not included.

How this calculator works

A small radioactive source that emits gamma rays behaves, from far enough away, like a point: its photons spread out evenly in every direction. That makes three things predictable. The dose rate is proportional to the activity (twice the curies, twice the dose rate). It falls inversely with the square of the distance. And each slab of shielding removes roughly a fixed fraction of what reaches it. Put together with a per-isotope constant, those rules give the dose rate anywhere around the source — or, run backwards, the activity that explains a meter reading, the distance to a boundary, or the thickness of lead, steel or concrete needed to reach a target.

Pick what you want to solve for, choose an isotope (or enter your own constant), and fill in the other values. Every input has its own unit selector, so you can mix curies and becquerels, meters and feet, or mR/h and µSv/h freely. The link button copies a URL that reproduces your exact inputs.

Gamma constants used

Values are exposure-rate constants for a point source, in R·m²/(Ci·h). Multiply by 10 for R·cm²/(mCi·h). Published tables disagree by 5–20 % depending on which x-rays and weak lines they include, so treat results as estimates. Some presets carry their own assumptions: Am-241 leaves out its low-energy neptunium x-rays, and Ra-226 assumes its short-lived daughters in equilibrium behind 0.5 mm of platinum. Each isotope's page and result note give its assumptions.

IsotopeΓ, R·m²/(Ci·h)Main photonsSource
Cs-1370.33661.7 keV (85.1 %, via Ba-137m)Unger, L. M. & Trubey, D. K.; U.S. Department of Health, Education, and Welfare; Cember, H. & Johnson, T. E.
Co-601.321173.2 keV (99.85 %), 1332.5 keV (99.98 %)Unger, L. M. & Trubey, D. K.; U.S. Department of Health, Education, and Welfare; Cember, H. & Johnson, T. E.
Ir-1920.48316.5 keV (82.8 %), 468.1 keV (47.8 %), 308.5 keV (30.0 %), 296.0 keV (28.7 %) and weaker linesUnger, L. M. & Trubey, D. K.; Cember, H. & Johnson, T. E.
Am-2410.01359.5 keV (35.9 %), plus Np L x-raysNational Nuclear Data Center, Brookhaven National Laboratory. NuDat 3; Hubbell, J. H. & Seltzer, S. M.
I-1310.22364.5 keV (81.5 %), 637.0 keV (7.2 %), 284.3 keV (6.1 %), 722.9 keV (1.8 %)Unger, L. M. & Trubey, D. K.; U.S. Department of Health, Education, and Welfare; Cember, H. & Johnson, T. E.
Tc-99m0.078140.5 keV (89 %), Tc K x-rays 18–21 keVUnger, L. M. & Trubey, D. K.; U.S. Department of Health, Education, and Welfare
Ra-2260.825Many lines from daughters Pb-214 and Bi-214 (e.g. 351.9 keV, 609.3 keV, 1764.5 keV)U.S. Department of Health, Education, and Welfare; Cember, H. & Johnson, T. E.
F-180.57511 keV annihilation (193.5 %)Unger, L. M. & Trubey, D. K.; U.S. Department of Health, Education, and Welfare
Na-221.21274.5 keV (99.94 %), 511 keV annihilation (180 %)Unger, L. M. & Trubey, D. K.; U.S. Department of Health, Education, and Welfare

Shielding data used

Broad-beam half-value and tenth-value layers (cm) from NCRP Report No. 49. They apply to the geometry and photon energies of those tables, and the calculator uses one constant layer for the whole thickness. Real transmission depends on geometry, spectrum and thickness, and first layers can differ from later ones. For a source with several photon lines or a thin shield, such as Ir-192, the spectrum hardens in the shield, so a constant layer can misstate the attenuation. Isotopes not listed have no broad-beam values in that table; for them, pick “Custom HVL” and enter a value from your own reference.

IsotopeLead HVL / TVLSteel (iron) HVL / TVLConcrete HVL / TVL
Cs-1370.65 / 2.1 cm1.6 / 5.3 cm4.8 / 15.7 cm
Co-601.2 / 4 cm2.1 / 6.9 cm6.2 / 20.6 cm
Ir-1920.6 / 2 cm1.3 / 4.3 cm4.3 / 14.7 cm
Ra-2261.66 / 5.5 cm2.2 / 7.4 cm6.9 / 23.4 cm

The formula

Ẋ = Γ · A · T / d²

T = 10^(−x / TVL)     or     T = 2^(−x / HVL)

A = Ẋ · d² / (Γ · T)          d = √(Γ · A · T / Ẋ)
x = TVL · log₁₀(Γ · A / (Ẋ · d²))

Ẋ is the exposure rate (R/h), Γ the specific gamma-ray constant (R·m²/(Ci·h)), A the activity (Ci), d the distance (m), x the shield thickness and T its transmission. Air-kerma rate is about Ẋ × 8.764 mGy/R (2.58 × 10⁻⁴ C/kg × 33.97 J/C, ICRU 85): the collision air kerma under electronic equilibrium, close to total air kerma at these energies, and not tissue dose. Dose-equivalent units use the photon convention 1 R ≈ 1 rem.

The point-source model holds once you are at least about three times the source's largest dimension away, measured from the center of the source. That depends on the source, so no single distance is safe to use for every source, and each result repeats the condition. It ignores air attenuation (small over a few meters, but substantial over tens to hundreds of meters), scatter from walls and floors, and self-shielding inside the source. It covers gamma rays only: it does not model neutrons, or the gamma rays produced when neutrons are captured in a shield, which need a separate shielding analysis.

Worked example

Portable gauge source rod. A nuclear moisture–density gauge typically holds about 8 mCi of Cs-137. With the rod extended and nothing in the way, at 1 m: Ẋ = 0.33 × 0.008 / 1² = 2.64 mR/h (roughly 26.4 µSv/h). Step back to 3 m and it falls ninefold to 0.293 mR/h. Put 2 cm of lead between you and the rod and the 1 m figure becomes 0.295 mR/h (transmission 11.2 %).

Inverse-square distance example. How far from a 100 Ci Ir-192 radiography source, with no collimator, does the idealized exposure rate drop to 2 mR/h? d = √(0.48 × 100 / 0.002) ≈ 155 m. This photon approximation illustrates the arithmetic; it is not a surveyed boundary or a demonstration of regulatory compliance. Air attenuation, scatter and actual source geometry matter at this distance. Radiographers use collimators, shielded exposure devices and measured surveys.

Frequently asked questions

What is a specific gamma-ray constant?
It is the exposure rate one unit of activity produces at one unit of distance from an unshielded point source, for example 0.33 R/h at 1 m from 1 Ci of Cs-137. It bundles the photon energies, their yields per decay and how strongly air absorbs them into one number per nuclide.
How much does the dose rate drop when I double the distance?
For a point source, with negligible attenuation and scatter, it falls to one quarter (the inverse-square law). Tripling the distance cuts it to one ninth. Distance is measured from the source itself, and close to large or extended sources the fall-off is slower. This is why distance is often the cheapest and most effective protection.
Why does the calculator give R/h and also mSv/h?
The equation produces exposure rate in roentgens per hour. Air-kerma rate (Gy/h) uses about 8.76 mGy of air kerma per R: the exact definition of the roentgen (2.58 × 10⁻⁴ C/kg) times the measured energy per unit charge in dry air (33.97 J/C), assuming electronic equilibrium and negligible radiative losses. That is a quantity in air, not the absorbed dose in tissue. For photons, radiation-protection practice treats 1 R as about 1 rem (10 mSv); that conversion is an approximation and is labeled as such.
What is the difference between HVL and TVL?
A half-value layer (HVL) is the thickness of shielding that halves the dose rate; a tenth-value layer (TVL) cuts it to one tenth. In the simple exponential model one TVL is about 3.3 HVLs; tabulated broad-beam values differ somewhat (for Cs-137 in lead, 2.1 cm against 0.65 cm). The broad-beam values used here include photons scattered inside the shield under the reference conditions of the NCRP 49 tables; other geometries and spectra can differ.
How much lead do I need to shield Cs-137?
With the broad-beam values used here (HVL 0.65 cm, TVL 2.1 cm of lead), each 2.1 cm of lead cuts the Cs-137 dose rate by about a factor of ten. Choose "Shield thickness" in the calculator for an estimate of the thickness for your target dose rate. It is an illustrative figure from a constant-layer model, not a shielding design.
Can I use this to decide whether an area is safe?
No. This is an educational estimate for a bare point source. Real sources have encapsulation, self-absorption and scatter from nearby surfaces. Use a calibrated survey meter and follow your radiation safety program.

Sources

  1. Unger, L. M. & Trubey, D. K. (1982). Specific Gamma-Ray Dose Constants for Nuclides Important to Dosimetry and Radiological Assessment. ORNL/RSIC-45/R1, Oak Ridge National Laboratory. www.osti.gov
  2. U.S. Department of Health, Education, and Welfare (1970). Radiological Health Handbook, revised edition. Public Health Service, Bureau of Radiological Health.
  3. Cember, H. & Johnson, T. E. (2009). Introduction to Health Physics, 4th edition. McGraw-Hill. (Chapters on external radiation safety and shielding.)
  4. NCRP Report No. 49 (1976). Structural Shielding Design and Evaluation for Medical Use of X Rays and Gamma Rays of Energies up to 10 MeV. National Council on Radiation Protection and Measurements. (Broad-beam half-value and tenth-value layers for gamma sources.)
  5. ICRU Report 85 (2011). Fundamental Quantities and Units for Ionizing Radiation. (Mean energy per ion pair in dry air, W/e = 33.97 J/C.)
  6. National Nuclear Data Center, Brookhaven National Laboratory. NuDat 3 (ENSDF evaluated half-lives and photon emission data). www.nndc.bnl.gov
  7. Hubbell, J. H. & Seltzer, S. M. (2004). Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients (version 1.4). NIST Standard Reference Database 126. www.nist.gov
  8. U.S. Nuclear Regulatory Commission. 10 CFR 20.1004, Units of radiation dose. www.nrc.gov